Measurement of long lived radioactive impurities retained in the disposable cassettes on the Tracerlab MX system during the production of [18F]FDG

Ferguson, D., Orr, P., Gillanders, J., Corrigan, G. and Marshall, Christopher 2011. Measurement of long lived radioactive impurities retained in the disposable cassettes on the Tracerlab MX system during the production of [18F]FDG. Applied Radiation and Isotopes 69 (10) , pp. 1479-1485. 10.1016/j.apradiso.2011.05.028

Full text not available from this repository.

Abstract

Using a High- Purity Germanium gamma-ray spectrometer, a number of radioisotopes have been identified within Tracerlab MX radiochemistry system cassettes used to synthesise [18F]FDG. Twenty radiochemistry cassettes were measured and the average total activity of each radioisotope was determined. Using these values and decay correction, the minimum time the cassettes should be left in a decay store before the specific activity falls below 0.4B q/g, the limit for disposal alongside Clinical Waste was found to be 24 months.