Development and evaluation of a novel protective device for upper gastrointestinal endoscopy in the COVID-19 pandemic: the EBOX

Mcleod, Robert William John, Warren, Neil and Roberts, Stuart Ashley 2021. Development and evaluation of a novel protective device for upper gastrointestinal endoscopy in the COVID-19 pandemic: the EBOX. Frontline Gastroenterology 12 (3) , pp. 188-192. 10.1136/flgastro-2020-101542

[thumbnail of Mcleod. Development and evaluation.pub.pdf]
Preview
PDF - Published Version
Available under License Creative Commons Attribution Non-commercial.

Download (756kB) | Preview

Abstract

Background During the COVID-19 pandemic, aerosol-generating procedures such as upper gastrointestinal endoscopy (UGIE) have been considered high risk. We designed a novel acrylic box (endoscopy box (EBOX)) with the intention of limiting aerosol and droplet spread during such procedures. We evaluated clinical utility, impact on the endoscopy team and also assessed the impact of the EBOX on macroscopic droplet spread from a simulated cough during UGIE. Methods Clinical utility was evaluated prospectively via EBOX use on 15 patients undergoing endoscopic retrograde cholangiopancreatography (13) or endoscopic ultrasound (2). Feedback was recorded from the endoscopy team regarding ease of positioning, impact of the EBOX on procedural performance and cleaning. A cough was simulated via explosion of a hyperinflated balloon containing 0.75 mL of ultraviolet disclosing lotion within the oral cavity of a mannequin, with and without the EBOX. Macroscopic spread was then evaluated with a ultraviolet torch. Results Three endoscopists and the team members found that the EBOX did not hamper the procedure and felt it was a useful adjunct to full personal protective equipment (PPE). Simulated cough without the EBOX identified macroscopic spread up to 2.3 m away from the patient’s mouth as well as onto key areas such as the exposed neck of the endoscopist, which is not considered in current PPE guidance. Simulated cough using the EBOX significantly reduced macroscopic spread onto key areas of the healthcare workers. Conclusions The EBOX is a valuable adjunct to recommended PPE for UGIE, but still allows these procedures to be performed in the standard manner.

Item Type: Article
Date Type: Publication
Status: Published
Schools: Psychology
Additional Information: This is an open access article distributed in accordance with the Creative Commons Attribution Non Commercial (CC BY-NC 4.0)
Publisher: BMJ Publishing Group
ISSN: 2041-4137
Date of First Compliant Deposit: 15 June 2020
Date of Acceptance: 7 June 2020
Last Modified: 04 May 2023 19:56
URI: https://orca.cardiff.ac.uk/id/eprint/132440

Citation Data

Cited 4 times in Scopus. View in Scopus. Powered By Scopus® Data

Actions (repository staff only)

Edit Item Edit Item

Downloads

Downloads per month over past year

View more statistics